1.Working condition
Operating environmental temperature: 18-35℃
Operating environmental humidity: ≤80%
Operating voltage: 220 V (±10%), 50Hz (±2%)
2.Single quadrupole mass spectrometer parameters
High-precision all-metal quadrupole equipped with pre-quadrupole;
Pre-quadrupole disassembly for cleaning to avoid contamination of the main quadrupole;
No temperature control is required to ensure quality stability indicators;
Optimizing ion energy technology;
Main performance indicators
Sensitivity: 1 pg, OFN S/N≥1500:1
Quality range: 1.5-1250 amu
Quality stability: ± 0.10 amu/48 hours
Quality accuracy: ± 0.10 amu
Maximum scanning speed: 20000 amu/s, with adjustable speed throughout the entire process
Resolution: 0.4-4amu adjustable
Dynamic range: 107
Peak area reproducibility: < 2% RSD
Scanning methods: full scan, selective ion monitoring, full scan and selective ion synchronous monitoring, alternating scanning.
3.Ion source
An inert ceramic ion source ensures that analytical substances will not adsorb on the ion source during long-term use, extending service life ;
Dual filament design, alternating use to extend filament replacement time;
Visual judgment of filament working status through the filament observation port;
Type of ion source: Electron impact source (EI)
Ion source configuration: Inert ceramic ion source, dual filament software switching
Filament current: 0-500 μA
Ionization energy: 5-100 eV
Ion source temperature: 50℃-350℃
Equipped with a filament observation window to visually judge the working status of the filament
Interface temperature: 50-350℃
4.Quality analyzer
Quality analyzer: equipped with high-precision all metal quadrupole with pre quadrupole
The quadrupole is optimized for temperature control, which can ensure quality stability indicators without the need for temperature control
Pre quadrupole detachable and washable to avoid contamination of the main quadrupole
The quadrupole has optimized ion energy technology, which optimizes the speed of ions passing through the quadrupole and improves signal quality
5.Detector
Long lifespan level 13 dispersed dynode electronic multiplier
Equipped with a 10kV conversion dynode and a lens system for removing neutral noise
6.Vacuum system
Front stage pump, speed≥4 m³/h
Configure a high-performance turbomolecular pump,speed≥250 L/s
Maximum allowable chromatographic column flow rate: 5 ml/min (He)
Configured with a hot cathode ionization vacuum gauge, and the software can directly read high vacuum readings.
7.Split/non split injector
Maximum using temperature:450˚C
Gas control mode: constant pressure, constant flow rate, program up/down pressure, program up/down flow, pulse injection
Program boosting/current raising:10th order
Pressure setting range:0-1035 kPa (0-150 psi)
Maximum split ratio:9999.9:1
Flow setting range: 0-200 mL/min (N2)
Flow setting range: 0-1000 mL/min (H2)
8.Column oven
Operating temperature range: room temperature: +4˚C to 450˚C
Temperature Setting Accuracy:0.1˚C
Temperature control accuracy:0.01˚C
Temperature Stability: ≤0.5%
Temperature uniformity: ≤2.5%
Environmental sensitivity: When the ambient temperature changes by 1˚C, better than 0.01℃
Heating Rate::≥100˚C /min
Programmed Heating Repeatability: ≤1%
Programmed Heating Order: 32
Cooling Time (from 450℃to 50℃): ≤4 min
Maximum Running Time: 9999.999 min