Reliable fully automatic graphite furnace analysis
Perfect safety protection measures
Advanced and reliable electronic design
Easy and practical analysis software
Main Specification |
Wavelength range | 190-900nm |
Wavelength accuracy | Better than ±0.25nm | |
Resolution | Two spectral lines of Mn at 279.5nm and 279.8nm can be separated with the spectral bandwidth of 0.2nm and valley-peak energy ratio less than 30%. | |
Baseline stability | 0.004A/30min | |
Background correction | The D2 lamp background correction capability at 1A is better than 30 times.The S-H background correction capability at 1.8A is better than 30 times. | |
Light Source System |
Lamp turret |
Motorized 6-lamp turret (Two high performance HCLs can be mounted on the turret to increase the sensitivity in flame analysis.) |
Lamp current adjustment | Wide pulse current: 0~25mA, Narrow pulse current: 0~10mA. | |
Lamp power supply mode | 400Hz square wave pulse;100Hz narrow square wave pulse + 400Hz wide square pulse wave. | |
Optical System |
Monochomator |
Single beam, Czerny-Turner design grating monochromator |
Grating |
1800 l/mm | |
Focal length |
277mm | |
Blazed Wavelength |
250nm | |
Spectral Bandwidth |
0.1nm, 0.2nm, 0.4nm, 1.2nm, auto switch over | |
Flame Atomizer |
Burner |
10cm single slot all-titanium burner |
Spray chamber |
Corrosion resistant all-plastic spray chamber. | |
Nebulizer |
High efficiency glass nebulizer with metal sleeve, sucking up rate: 6-7mL/min | |
Emission burner provided | ||
Graphite Furnace |
Temperature range |
Room temperature~3000ºC |
Heating rate |
2000℃/s | |
Graphite tube dimensions | 28mm (L) x 8mm (OD) | |
Characteristic mass |
Cd≤0.8 ×10-12g, Cu≤5 ×10-12g, Mo≤1×10-11g | |
Precision |
Cd≤3%, Cu≤3%, Mo≤4% | |
Detection and Data Processing System |
Detector |
R928 photomultiplier with high sensitivity and wide spectral range. |
Software |
Under Windows operating system | |
Analytical method |
Working curve auto-fitting; standard addition method; automatic sensitivity correction; automatic calculation of concentration and content. | |
Repeat times | 1~99 times, automatic calculation of mean value, standard deviation and relative standard deviation. | |
Multi-task Functions |
Sequential determination of multi-elements in the same sample. | |
Condition reading |
With model function | |
Result printing |
Measurement data and final analytical report printout, editing with Excel. | |
Standard RS-232 serial port communication | ||
Graphite Furnace Autosampler | Sample tray capacity | 55 sample vessels and 5 reagent vessels |
Vessel material |
Polypropylene | |
Vessel volume |
3ml for sample vessel, 20ml for reagent vessel | |
Minimum sampling volume |
1μl | |
Repeatable sampling times |
1~99 times | |
Sampling system |
Accurate dual pump system, with 100μl and 1ml injectors. | |
Characteristic Concentration and Detection Limit |
Air-C2H2 flame |
Cu: Characteristic concentration ≤ 0.025 mg/L, Detection limit≤0.006mg/L; |
Function Expansion |
Hydride vapor generator can be connected for hydride analysis. | |
Dimensions and Weight |
Main unit |
107X49x58cm, 140kg |
Graphite furnace |
42X42X46cm, 65kg | |
Autosampler |
40X29X29cm, 15kg |