• head_banner_01

WFX-320 Flame Atomic Absorption Spectrophotometer

Short Description:


Product Detail

Product Tags

Features

High cost-effective flame AAS

Reasonable design, adopting the same key parts as in high end instruments, ensures basic functions but less automation to provide an economic model for users.

Reliable integration of main unit with microprocessor

Built-in microprocessor with necessary auto-control and data processing functions achieve high reliability of the instrument.

Simple and easy operation

Eye-catching digital display, multi-function data processing ability and fast function-key direct input realize easy and fast analysis.

Specifications

Main Specifications Wavelength range 190-900nm
Wavelength accuracy 0.5nm
Resolution Two spectral lines of Mn at 279.5nm and 279.8nm can be separated with the spectral bandwidth of 0.2nm and valley- peak energy ratio less than 30%
Baseline stability 0.005A/30min
Background correction The D2 lamp background correction capability at 1A is better than 30 times
Light Source System 2 lamps are powered simultaneously (one preheating)
Lamp current adjustment range: 0-20mA
Lamp power supply mode Powered by 400Hz square pulse
Optical System Monochromator Single beam, Czerny-Turner design grating monochromator
Grating 1800 I/mm
Focal length 277mm
Blazed wavelength 250nm
Spectral bandwidth 0.1 nm, 0.2nm, 0.4nm, 1.2nm 4 steps
Adjustment Manual adjustment for wavelength and slit
Flame Atomizer Burner 10cm single slot all-titanium burner
Spray chamber Corrosion resistant all-plastic spray chamber
Nebulizer High efficiency glass nebulizer with metal sleeve, sucking up rate: 6-7ml/min
Position adjustment Manual adjusting mechanism for vertical, horizontal positions and the rotation angle of the burner
Gas line protection Fuel gas leakage alarm
Detection and Data Processing System Detector R928 Photomultiplier with high sensitivity and wide spectral range
Electronic and micro-computer system Automatic adjustment of light source power. Light energy and negative high-voltage auto-balance
Display mode LED display of energy and measurement values, concentration direct reading
Read mode Transient, time average, peak height, peak area Integral time is selectable in the range of 0.1-19.9s.
Scale expansion 0.1-99
Data processing mode Automatic calculation of mean, standard deviation and relative standard deviation. Repeating number is in the range of 1-99
Measurement mode Automatic curve fitting with 3-7 standards; Sensitivity auto-correction
Result printing Measurement data, working curve, signal profile and analytical conditions can all be printed out.
Instrument self-check Check status of each function key
Characteristic Concentration and Detection Limit Air-C2H2 flame Cu: Characteristic concentration≦ 0.025mg/L, Detection limit ≦ 0.006mg/L;
Function Expansion Hydride vapor generator can be connected for hydride analysis
Dimensions and Weight 1020x490x540mm,80kg unpacked

  • Previous:
  • Next:

  • Write your message here and send it to us